MS2306

MS2306

MS2115A

MS2115A

MS2308

US$1,797.95

Feature

● Digital / Analog dual display 9999 counts.
● Auto Power Off.
● Clock.
● AVG.
● Compensation Resistance.
● Alarming indicator for limit-value measurement.
● Battery capacity indicator.
● Display Backlight.
● Setting length between resistivity testing point.
● Data Storage 100 groups.

DC Voltage AC Voltage Resistance
Current
Current
AC -
DC -
Voltage
Voltage
AC 48V
DC 20V
SKU
MS2308CTGLO

MASTECH MS2308 Advanced Earth Resistance Tester: Reliable and Comprehensive Ground Resistance Measurement Tool

The MASTECH MS2308 is a versatile earth ground resistance tester designed for accurate and reliable measurement of ground resistance and soil resistivity. It features both digital and analog displays, making it ideal for professional use in electrical and construction fields where accuracy ground resistance measurements are essential.

Key Features

  • Multifunctional Measurement: including 3-pole and 4-pole earth resistance testing, 2-pole resistance, soil resistivity measurement, compensation for cable resistance, and Interfering Voltage/Current/Frequency.
  • Max Display: 9999 counts for clear and detailed readings.
  • Dual Display: Features both digital and analog displays for real-time monitoring and accurate data representation.
  • Data Hold: Allows users to freeze the displayed measurement for easy recording and analysis.
  • Data Logging: Stores up to 100 groups of measurement data for later review.
  • Backlight: Ensures readability in low-light environments.
  • USB Interface: Allows for data transfer and analysis on a computer.
  • Additional Features: Includes clock, alarm indication for limit value measurements, auto power off, and battery capacity Indicator.

Specifications

  • 3 Pole & 4 Pole Earthing-Resistance Test: 0.02Ω~300kΩ
    • Test Voltage: AC 20/48V
    • Test Frequency: 94Hz/128Hz
  • Selective mode (3 pole clamp /4 pole clamp): 0.02Ω~20kΩ
    • Test Voltage: AC 20/48V
    • Test Frequency: 94Hz/128Hz
  • Non-auxiliary-polar mode (clamp): 0.02Ω~150kΩ
    • Test Voltage: AC 20/48V
    • Test Frequency: 94Hz/128Hz
  • 2 Pole R (AC): 0.02Ω~300kΩ
    • Test Voltage: AC 20V/48V
    • Test Frequency: 94Hz/128Hz
  • 2/4 Pole R (DC): 0.02Ω~3kΩ
    • Interfering Voltage: 1~50V DC/AC
    • Interfering Frequency: 16Hz~400Hz
  • Soil Resistivity ρ: 0.02~1000
    • Test Voltage: AC 20/48V
    • Test Frequency: 94Hz/105Hz/111Hz/128Hz/AFC
  • RK (Wire Compensation): 30Ω
    • Test Voltage: AC20/48V
    • Test Frequency: 94Hz/105Hz/111Hz/128Hz/AFC
  • Interfering Voltage/Current/Frequency
    • Interfering Voltage: 1~50V DC/AC Test
    • Interfering Frequency: 16Hz~400Hz
    • Interfering Current: 20mA~2A
  • Short Circuit Current ≤250mA

Benefits

  • Versatility: Supports 3-pole, 4-pole, and non-auxiliary-polar testing modes.
  • Portability: Compact design for on-site testing.
  • Safety: Enhanced with multiple certifications.
  • Data Management: USB interface for easy data transfer.

Why Choose MASTECH MS2308?

Choose the MS2308 for its comprehensive range of measurement options, durable design, and advanced features like data storage and USB connectivity. Its versatile testing capabilities make it a valuable tool for any professional needing accurate earth resistance and soil resistivity measurements.

Information
Description Value
Display counts 9999
DC Voltage 20V
AC Voltage 48V
3 Pole & 4 Pole Resistance No
Resistance 20kΩ
Data Hold No
Resistivity of Soil No
Overrange Indication Yes
Low Battery Indicator Yes
Backlight Yes
Data Logging 100 groups
Interface USB

Methods

Specifications

Resolution

Accuracy

Range

3 Pole & 4 Pole
Earthing-Resistance Test

Test Voltage: AC20/48V

Test Frequency: 94Hz/128Hz

Earthing-Resistance Test

0.001Ω


±(5%+10)


0.02Ω~300kΩ

Selective mode
(3 pole clamp /4 pole clamp)

Test Voltage: AC20V

Test Frequency: 94Hz/128Hz

Earthing-Resistance Test

0.001Ω

±(5%+10)

0.02Ω~20kΩ

Non-auxiliary-polar mode
( clamp )

Test Voltage: AC20/48V

Test Frequency: 94Hz/128Hz

Range: (kΩ・m)

0.001Ω

±(10%+10)

0.02Ω~150Ω

2 Pole R (AC)

Test Voltage: AC20/48V

Test Frequency: 94Hz/128Hz

Range:

0.001Ω

±(5%+10)

0.02Ω~300kΩ

2/4 Pole R (DC)

Interfering Voltage: 1~50V DC/AC

Interfering Frequency: 16Hz~400Hz

0.001Ω

±(5%+10)

0.02Ω~3kΩ

Soil Resistivity ρ

Test Voltage: AC20/48V

Test Frequency: 94Hz/105Hz/111Hz/128Hz/AFC

Range: (kΩ・m)

0.001Ω

±(5%+10)

0.02~1000

RK
(Wire Compensation)

Test Voltage: AC20/48V

Test Frequency: 94Hz/105Hz/111Hz/128Hz/AFC

Range:

0.001Ω

±(3%+10)

30Ω

Interfering Voltage/
Current/Frequency

Interfering Voltage: 1~50V DC/AC

Interfering Frequency: 16Hz~400Hz

Interfering Current: 20mA~2A

0.1V

1Hz

1mA

±(5%+5)

±(1%+10)

±(5%+5)

 

Short Circuit Current ≤ 250mA

  • Industrial Use
  • Soil Resistivity Measurement
  • Grounding System Check
General
Description Value
Power Supply (a)6XLR14 alkaline(2#); rated voltage: 6x1.5V, (b)Recharge Battery: NI~MH Recharge Battery, 7.2V, (c)Charger: Input voltage: 100V~240V; Frequency: 50Hz~60Hz; Output Voltage: 12VDC 3A
Product Size 284mmx230mmx125mm/11.2"x9.1"x4.9"
Accessories Test wire, Alligator clip, Auxiliary earth bar, Power Cord, Current Transformer φ51mm, Current clamp cable, USB interface cable, Software CD, Carry bag, Rechargeable Ni-MH battery (optional), Charger (Optional), Battery and user's manual
Carton Box 520mmx330mmx220mm
Carton included 1